Volume 30 Issue 6
Jun.  2018
Turn off MathJax
Article Contents
Wang Jianshu, Zheng Kun, Wang Honggang, et al. Error analysis of microwave power measurement instruments[J]. High Power Laser and Particle Beams, 2018, 30: 063006. doi: 10.11884/HPLPB201830.170455
Citation: Wang Jianshu, Zheng Kun, Wang Honggang, et al. Error analysis of microwave power measurement instruments[J]. High Power Laser and Particle Beams, 2018, 30: 063006. doi: 10.11884/HPLPB201830.170455

Error analysis of microwave power measurement instruments

doi: 10.11884/HPLPB201830.170455
  • Received Date: 2017-11-10
  • Rev Recd Date: 2018-02-24
  • Publish Date: 2018-06-15
  • Power meters, oscilloscopes, detectors, and vector network analyzers are commonly used in microwave measurements. The introduced errors by these instruments have direct impact on the measurement accuracy. In this paper, the measurement methods of power meters, oscilloscopes, detectors, and network analyzers are described, based on which the introduced errors by the typical measurement instruments under different measurement conditions are analyzed. The experiment results demonstrate that the measured power by the detectors is at most 0.4 dB bigger than that by the oscilloscope. The maximum difference of the measured power by the oscilloscope under different suppression bandwidths is 0.3 dB. The measured power by the vector network analyzer N9917A is 1 dB bigger than that by the analyzer AV3672 at 4 GHz with the attenuation of 60 dB.
  • loading
  • [1]
    舒挺, 王勇, 李继建, 等. 高功率微波的远场测量[J]. 强激光与粒子束, 2003, 15(5): 142-145. http://www.hplpb.com.cn/article/id/137

    Shu Ting, Wang Yong, Li Jijian, et al. Measurement of high power microwaves in the far-field zones. High Power Laser and Particle Beams, 2003, 15(5): 142-145 http://www.hplpb.com.cn/article/id/137
    [2]
    王宏军, 黄文华, 刘国治. 几种高功率微波频率诊断方法[J]. 微波学报, 2001, 17(1): 62-66. https://www.cnki.com.cn/Article/CJFDTOTAL-WBXB200101012.htm

    Wang Hongjun, Huang Wenhua, Liu Guozhi. Several methods for high power microwave frequency diagnosis. Journal of Microwaves, 2001, 17(1): 62-66 https://www.cnki.com.cn/Article/CJFDTOTAL-WBXB200101012.htm
    [3]
    汤世贤. 微波测量[M]. 北京: 国防工业出版社, 1991.

    Tang Shixian. Measurement of microwaves. Beijing: National Defense Industry Press, 1991
    [4]
    Kancleris Z. Resistive sensor for short HPM pulse measurement[R]. Vilnius: Semiconductor Physics Institute, 1998.
    [5]
    张黎军, 陈昌华, 腾雁, 等. 高功率微波常用测量器件标定方法研究[J]. 微波学报, 2016, 32(5): 60-64. https://www.cnki.com.cn/Article/CJFDTOTAL-WBXB201603014.htm

    Zhang Lijun, Chen Changhua, Ten Yan, et al. Research on the calibration methods of HPM measuring elements. Journal of Microwavs, 2016, 32(5): 60-64 https://www.cnki.com.cn/Article/CJFDTOTAL-WBXB201603014.htm
    [6]
    李松. 毫米波检波器研制[D]. 成都: 电子科技大学, 2008.

    Li Song. The investigation on millimeter microwave detector. Chengdu: University of Electronic Science and Technology of China, 2008
    [7]
    晏峰, 崔新红, 李治国, 等. X波段高功率微波辐射场测量角锥天线特性[J]. 强激光与粒子束, 2011, 23(11): 3145-3148. http://www.hplpb.com.cn/article/id/5509

    Yan Feng, Cui Xinhong, Li Zhiguo, et al. Characteristics of pyramidal horn used in HPM radiation field measurement at X-band. High Power Laser and Particle Beams, 2011, 23(11): 3145-3148 http://www.hplpb.com.cn/article/id/5509
    [8]
    方进勇, 张治强, 黄文华, 等. 常用衰减器功率容量测试[J]. 强激光与粒子束, 2010, 22(11): 2663-2668. http://www.hplpb.com.cn/article/id/4872

    Fang Jinyong, Zhang Zhiqiang, Huang Wenhua, et al. Power capacity test of common attenuator. High Power Laser and Particle Beams, 2010, 22(11): 2663-2668 http://www.hplpb.com.cn/article/id/4872
    [9]
    刘英君, 宁辉, 周捷, 等. 波导器件网络参数的误差修正方法研究[J]. 电子测量与仪器学报, 2013, 27(5): 467-472. https://www.cnki.com.cn/Article/CJFDTOTAL-DZIY201305015.htm

    Liu Yingjun, Ning Hui, Zhou Jie, et al. Research on error correction method for network parameter of waveguide device. Journal of Electronic Measurement and Instrument, 2013, 27(5): 467-472 https://www.cnki.com.cn/Article/CJFDTOTAL-DZIY201305015.htm
    [10]
    刘英君. 波导同轴转换器件误差修正模型的研究[D]. 西安: 西北核技术研究所, 2011.

    Liu Yingjun. Research on the error correction model of waveguide-coaxial devices. Xi'an: Northwest Institute of Nuclear Technology, 2011
  • 加载中