tang dao-run, jiang shao-en, wu deng-xue, et al. X-ray calorimeters used for measurement in laser-fusion experiments[J]. High Power Laser and Particle Beams, 2005, 17.
Citation: tang dao-run, jiang shao-en, wu deng-xue, et al. X-ray calorimeters used for measurement in laser-fusion experiments[J]. High Power Laser and Particle Beams, 2005, 17.

X-ray calorimeters used for measurement in laser-fusion experiments

  • Publish Date: 2005-05-15
  • X-ray calorimeters are ready to measure the total soft X-ray energy emitted from the plasma produced by laser because of their bodily absorption, linear response, insensitivity to the electromagnetic disturbance, and so on. The calorimeters mainly include absorbers, thermocouples, bases and shrouds. When X-rays are deposited in the absorders, photon energy absorbed is quickly converted into intrinsic energy which simultaneously dissipates by thermal conduction and radiation. The X-ray calorimeters were absolutely on-line calibrated in Shenguang-Ⅱ laser facility with the X-ray diode array spectrometer which has been absolutely calibrated on Beijing Synchrotron Radiation Facility. 20 shots' experimental results show that the X-ray calorimeters are stable, the sensitivity of calorimeter is (