Volume 20 Issue 01
Jan.  2008
Turn off MathJax
Article Contents
ni wei-feng, xu jia-yun, chen zhong, et al. Calibration in measurement of ICF target material areal density using β-ray[J]. High Power Laser and Particle Beams, 2008, 20.
Citation: ni wei-feng, xu jia-yun, chen zhong, et al. Calibration in measurement of ICF target material areal density using β-ray[J]. High Power Laser and Particle Beams, 2008, 20.

Calibration in measurement of ICF target material areal density using β-ray

  • Publish Date: 2008-01-15
  • ICF target material areal density and its distribution can be measured by counting beta particles passing through the material. It is necessary to calibrate the areal density measuring device using a series of standard materials with known areal densities and the same components with the materials to be analyzed. For the density measurement of a kind of ICF target material the calibrating method is proposed. This calibrating method uses the approximate, materials with composition of the ICF target material as the standard material. The transmissivity of beta particles possessing different energy can be simulated and computed by Monte-Carlo method when they pass different materials. The corresponding error to the measuring result is also obtained. The error is only several percent using pap
  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索
    Article views (2340) PDF downloads(565) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return