Qian Yikun, Wei Biao, Liu Yixin, et al. Calibration method and application of portable gamma ray dose(rate) meters based on minitype reference radiationJ. High Power Laser and Particle Beams, 2018, 30(9): 096007. DOI: 10.11884/HPLPB201830.180091
Citation: Qian Yikun, Wei Biao, Liu Yixin, et al. Calibration method and application of portable gamma ray dose(rate) meters based on minitype reference radiationJ. High Power Laser and Particle Beams, 2018, 30(9): 096007. DOI: 10.11884/HPLPB201830.180091

Calibration method and application of portable gamma ray dose(rate) meters based on minitype reference radiation

  • The existing calibration methods of gamma ray dose (rate) meters have the disadvantages of long distance, long period and low efficiency. To overcome these disadvantages, the Institute of Nuclear Physics and Chemistry of the Chinese Academy of Engineering Physics applies a machine learning prediction method using sample instruments to carry out a research on field calibration technology for portable gamma dose (rate) meters based on minitype reference radiation. A field calibration device for gamma dose (rate) meters for radiation protection was successfully developed, which achieved a satisfactory result with a measurement standard uncertainty of no more than 5%. This paper also presents a new method for determination of the conventional true value of gamma ray air kerma, and provides a novel model equipment for field calibration of the portable gamma dose (rate) meters.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return