Liu Qinghua, Li Jing, Shan Lijun, et al. High-energy CT system with 10 lp/mm spatial resolutionJ. High Power Laser and Particle Beams, 2022, 34(12): 124001. DOI: 10.11884/HPLPB202234.220322
Citation: Liu Qinghua, Li Jing, Shan Lijun, et al. High-energy CT system with 10 lp/mm spatial resolutionJ. High Power Laser and Particle Beams, 2022, 34(12): 124001. DOI: 10.11884/HPLPB202234.220322
  • Based on the photocathode S-band electron linac, a high-energy (9 MeV) micro-focus CT system named “Jingwei” was built. The focal-point is less than 100 μm, and the 7 h dose fluctuation is as low as 1.6%. An initial imaging experiment was also carried out. The focal spot of the dual-wire image quality meter clearly distinguished the 13D wire with a wire diameter of 0.05 mm, and the CT spatial resolution measured by a test card is better than 10 lp/mm. The electron beam energy of the system is also adjustable from 6 MeV to 18 MeV.