wei biao, ren yong, tang yuelin, et al. Novel method for measuring and correcting dead-time using 252Cf neutron source based autocorrelation function in nuclear radiation detectionJ. High Power Laser and Partical Beams, 2010, 22(10).
Citation: wei biao, ren yong, tang yuelin, et al. Novel method for measuring and correcting dead-time using 252Cf neutron source based autocorrelation function in nuclear radiation detectionJ. High Power Laser and Partical Beams, 2010, 22(10).

Novel method for measuring and correcting dead-time using 252Cf neutron source based autocorrelation function in nuclear radiation detection

  • To solve the problem of dead time loss, a novel method for measuring and correcting dead-time is proposed with the use of autocorrelation function based on a 252Cf -source-driven spectrum analyzing system. The experiment shows that the dead-time of the analyzing system is 40 ns. The dead-time can be obtained directly from the spectrum of autocorrelation with high measuring precision of nanosecond level.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return